Advanced Search
CAO Zheng-lin, SHEN Jian-xin, LIAO Wen-he. Research of excimer laser corneal surgery and flying-spot scanning algorithm[J]. LASER TECHNOLOGY, 2006, 30(6): 631-635.
Citation: CAO Zheng-lin, SHEN Jian-xin, LIAO Wen-he. Research of excimer laser corneal surgery and flying-spot scanning algorithm[J]. LASER TECHNOLOGY, 2006, 30(6): 631-635.

Research of excimer laser corneal surgery and flying-spot scanning algorithm

More Information
  • Received Date: November 06, 2005
  • Revised Date: December 30, 2005
  • Published Date: November 24, 2006
  • In order to increase the precision and safety of the corneal ablation refractive surgery,two mathematical models for corneal ablation were given,one was for spherical myopia and compound myopic astigmatism,the other was for spherical hyperopia and compound hyperopic astigmatism,Based on both the mathematical models,the flying-spot scanning algorithm was put forward.The mathematical models and the algorithm were applied to ophthalmic excimer laser system.Eighty-nine eyes underwent laser in situ keratomileusis(LASIK) using the models and algorithm,including 88 eyes with spherical myopia and compound myopic astigmatism and 1 with compound hyperopic astigmatism.The lightsight of 80.7% of the nake eyes increased upto 0.8.It can be concluded that the mathematical models and the algorithm are precise and safe and the laser cornea surgery is satisfactory.
  • [1]
    FARAH S,AZAR D,WONG J et al.Laser in situ keratomileusis:literature review of a developing technique[J].Journal of Cataract Refractive Surgery,1998,24(6):989~1006.
    [2]
    SALAH T,WARING G O,MAGHRABYA E L et al.Excimer laser in situ keratomileusis under a corneal flap for myopia of 2 to 20 diapters[J].America Journal of Ophthalmology,1996,121 (3):143~155.
    [3]
    ZHANG Y H,SHEN J X,LIAO W H.Use of excimer to correct compound myopic astigmatism[J].Optics and Precision Engineering,2003,11(3):291~295(in Chinese).
    [4]
    ZHANG Y H,SHEN J X,HU L G et al.Mathematical models for laser in situ keratomileusis and photorefractive keratectomy[J].Chinese Journal of Biomedical Engineering,2003,22 (4):289~295 (in Chinese).
    [5]
    ARTURO S,MAGALLANES R,MONTES M et al.Laser in situ keratomileusis for simple myopic,mixed,and simple hyperopic astigmatism[J].Journal of Refractive Surgery,1998,14(2):175~177.
    [6]
    LU W X.Excimer laser refractive corneal surgery[M].Beijing:Scientific and Technical Documents Publishing House,2000.46~52.(in Chinese).
    [7]
    LIN Zh N,SHEN J X,LIAO W H.Research of ablation of 193 nm excimer laser on cornca[J].Laser Technology,2003,27(6):510~513(in Chinese).
    [8]
    TROKEL S L.History and mechanism of action of excimer laser corneal surgery[A].Corneal Laser Surgery[C].St Louis:Mosby Year Book Inc,1995.1~10.
    [9]
    ZHANG Zh X.Laser-tissue interactions:fundamentals and applications[M].Xi'an:Xi'an Jiaotong University Publishing House,1999.68~81 (in Chinese).