Advanced Search

ISSN1001-3806 CN51-1125/TN Map

Volume 34 Issue 6
Mar.  2011
Article Contents
Turn off MathJax

Citation:

Effect of residual chirp on ultrashort pulse width measured with autocorrelation method

  • Corresponding author: ZHU Qi-hua, qihzh@163.com
  • Received Date: 2009-11-02
    Accepted Date: 2010-05-05
  • To study the effect of residual chirp on ultrashort pulse width measurement by means of the second harmonic autocorrelation,starting from its measurement principle,the effect was simulated for ultrashort laser pulses with different chirps at center wavelength of 800nm and pulse width of 100fs.The results show that the measured value would be larger than factual value with positive chirp,smaller with negative chirp.Moreover,the measurement error will be augmented with the chirp.The effect of the thickness of β-BaB2O4 (BBO) crystal on the pulse width measurement was also discussed numerically.It is found that choosing thin nonlinear crystal could control the measurement error effectively.Using the nonlinear crystal width of L≤0.5Ld, the measurement error caused by chirp can be controlled to an ideal range.
  • 加载中
  • [1]

    MORGNER U,KARTNET F X.Sub-two-cycle pulses from a Kerr-lens mode-locked Ti: sapphire laser[J].Opt Lett,1999,24(6):411-413.
    [2]

    PENG H S.Ultraintense solid-state lasers and applications to the frontiers of sciences(1)[J].Chinese Journal of Lasers,2006,33(6):721-729(in Chinese).
    [3]

    PENG H S.Ultraintense solid-state lasers and applications to the frontiers of sciences(2)[J].Chinese Journal of Lasers,2006,33(7):865-873(in Chinese).
    [4]

    HAO X,ZHU Q H,WANG X.The influence to the single-shot second-order autocorrelator due to spatial irregular intensity distribution[J].Chinese Journal of Lasers,2008,35(10):1553-1557(in Chinese).
    [5]

    CHAI L,GAO F.Advances in the diagnostic techniques of femtosecond optical pulses[J].Journal of Optoelectronics·Laser,2002,13(6):647-651(in Chinese).
    [6]

    CHEN B F,LIU T F.Numerical simulation of femtosecond pulses measurement with SPIDER[J].Laser Technology,2006,30(1):15-17(in Chinese).
    [7]

    RAGHURAMAIAH M,SHARMA A K.A second-order autocorrelator for single-shot measurement of femtosecond laser pulse durations[J].Sadhana,2001,26(6):603-611.
    [8]

    OUYANG X P,ZHANG F L.Second order autocorrelation applied to femtosecond laser[J].Chinese Journal of Lasers,2009,36(3):742-745(in Chinese).
    [9]

    LAN X J.Laser technology[M].Beijing:Science Press,2000:169-171(in Chinese).
    [10]

    SMITH A,ARMSTRONG D J,ALFORD W J.Increased acceptance bandwidths in optical frequency conversion by use of multiple walk-off-compensating nonlinear crystals[J].J O S A,1998,B15(1):122-141.
    [11]

    LÜ T Z,WANG T,QIAN L J.Calculation of frequency doubling efficiency of femto second pulses in BBO crystals[J].Acta Physica Sinica,2002,51(6):1268-1272(in Chinese).
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article views(5031) PDF downloads(228) Cited by()

Proportional views

Effect of residual chirp on ultrashort pulse width measured with autocorrelation method

    Corresponding author: ZHU Qi-hua, qihzh@163.com
  • 1. Research Center of Laser Fusion, China Academy of Engineering Physics, Mianyang 621900, China;
  • 2. Graduate School, China Academy of Engineering Physics, Mianyang 621900, China

Abstract: To study the effect of residual chirp on ultrashort pulse width measurement by means of the second harmonic autocorrelation,starting from its measurement principle,the effect was simulated for ultrashort laser pulses with different chirps at center wavelength of 800nm and pulse width of 100fs.The results show that the measured value would be larger than factual value with positive chirp,smaller with negative chirp.Moreover,the measurement error will be augmented with the chirp.The effect of the thickness of β-BaB2O4 (BBO) crystal on the pulse width measurement was also discussed numerically.It is found that choosing thin nonlinear crystal could control the measurement error effectively.Using the nonlinear crystal width of L≤0.5Ld, the measurement error caused by chirp can be controlled to an ideal range.

Reference (11)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return