  • 摘要: 在对红外探测器进行理论分析的基础上,设计并研制了液氮温度下的Yba2Cu3-xZnxO7薄膜红外探测器,系统地测试了器件的特征参数.最好的结果为:对于波长为10μm,调制频率为f=500Hz,带宽为Δf=1Hz的红外输入辐射Rv(500,10,1)=3587V/W,NEP(500,10,1)=6.5×10-12W/Hz1/2,D*(500,10,1)=7.2×1012cmHz1/2/W,τ(500,10,1)=1.2ms.

     

    Abstract: Base on the theoretical analysis,the YBa2Cu3-xZnxO7 thin film infrared detector has been designed and manufactured.The properties of the detector has been measured at 77°K and are well agreement with design,as follow:Rv(500,10,1)=3587V/W,NEP(500,10,1)=6.5×10-12W/Hz1/2,D*(500,10,1)=7.2×1012cmHz1/2/W,τ(500,10,1)=1.2ms.

     

/