A common-path heterodyne interferometer technique for measurement of optical activity of liquid

  • Received Date: 2011-06-27
    Accepted Date: 2011-07-04
  • In order to improve the detection sensitivity of optical activity,based on Zeeman effect,a common-path heterodyne interferometer was developed to measure the optical activity of liquid.In this heterodyne interferometer,a pair of correlated orthogonal linear polarized beam is emitted from a Zeeman He-Ne laser,which reduces the complexity of the optical system greatly.A balanced detector and a differential amplifier are integrated into this interferometer so as to reduce the excess noise induced by laser intensity fluctuation and external environments significantly.Thus,a shot-noise-limited detection of this interferometer can be achieved.The optical activity of glucose solution with concentration of(0~50)mg/dL was measured with this common-path heterodyne interferometer and the detection sensitivity was up to 2.3×10-5 degree.This simple heterodyne interferometer can be applied in the optical activity analysis of solid-state,liquid-state or gas-liquid material.
  • 加载中
  • [1]

    LID T,LIU Y H.Study on the distinguished methods the optically active compounds[J].Theory and Practice of Contemporary Education,2011,3(3):50-51 (in Chinese).
    [2]

    LIU Z Q.Measuring specific rotation and concentration of the sucrose solution with light intensity equipment[J].College Physics,2010,29(2):37-42(in Chinese).
    [3]

    HUANG S J.Measure to revolve angular at solution of the sugar[J].Metrology & Measurement Technique,2009,36(4):88-89(in Chinese).
    [4]

    YARIY A,LOTSPEICH J F.Coupled-mode analysis of light propagation in optical active crystals[J].Journal of the Optical Society of America,1982,72(2):273-277.
    [5]

    SILVER M P,BADOZ J,BRIAT B.Chiral reflection from a naturally optically active medium[J].Optics Letters,1992,17 (12):886-888.
    [6]

    CHEUNG C K,NAIK P S,BELING C D,et al.Performance of a slow positron beam using a hybrid lens design[J].Applied Surface Science,2006,252(9):3132-3137.
    [7]

    MITSUI T,SAKURAI K.Precise measurement of the refractive index and optirotatory power of a suspension by a delayed optical heterodyne technique[J].Applied Optics,1996,35(13):2253-2258.
    [8]

    BARRON L.Molecular light scattering and optical activity[M].New York:Cambridge University Press,2004:386-388.
    [9]

    COTE D,VITAINI A.Balanced detection for low-noise precision polarimetric measurement of optically active[J].Journal of Biomedical Optics,2004,9(1):213-220.
    [10]

    ZHAO PT,LIGH,PENGHD,et al.New design of optical ectivity measurement system[J].Laser Technology,2005,29(5):504-506 (in Chinese).
    [11]

    FENG W W,SONG L K,CHEN L G.The study of the relation between circular polarized light' s refractive rate and wavelength[J].Laser Technology,2004,28(6):639-644(in Chinese).
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article views(4692) PDF downloads(702) Cited by()

Proportional views

A common-path heterodyne interferometer technique for measurement of optical activity of liquid

  • 1. Department of Physics, Shaanxi University of Technology, Hanzhong 723003, China

Abstract: In order to improve the detection sensitivity of optical activity,based on Zeeman effect,a common-path heterodyne interferometer was developed to measure the optical activity of liquid.In this heterodyne interferometer,a pair of correlated orthogonal linear polarized beam is emitted from a Zeeman He-Ne laser,which reduces the complexity of the optical system greatly.A balanced detector and a differential amplifier are integrated into this interferometer so as to reduce the excess noise induced by laser intensity fluctuation and external environments significantly.Thus,a shot-noise-limited detection of this interferometer can be achieved.The optical activity of glucose solution with concentration of(0~50)mg/dL was measured with this common-path heterodyne interferometer and the detection sensitivity was up to 2.3×10-5 degree.This simple heterodyne interferometer can be applied in the optical activity analysis of solid-state,liquid-state or gas-liquid material.