Advanced Search

ISSN1001-3806 CN51-1125/TN Map

Volume 28 Issue 6
Sep.  2013
Article Contents
Turn off MathJax

Citation:

Temperature measurement sensitivity of a practical real-time temperature measurement system

  • Received Date: 2003-11-20
    Accepted Date: 2004-04-17
  • Based on the Kirchhoff law,a practical temperature real-time measurement system using laser diode InGaAs/I and LiTaO3 pyroelectric detector is presented.By means of analyzing the relationships among resolving power of A/D converter,temperature measurement sensitivity of V(T)-T curve and temperature measurement range,16bite A/D converter for this instrument is determined.According to the relationships among temperature measurement accuracy,relative temperature measurement sensitivity of V(T)-T curve and waveleng ths,further optimal wavelength is discussed.Based on the relationships among resolving power of the detector and A/D converter and temperature sensitivity of V(T)-T curve,further optimal wavelength bandwidth is analyzed.Temperature measurement sensitivity of the instrument at 673K~1473K is given.Temperature measurement sensitivity is above 0.5K at lower temperatures and above 0.1K at higher temperatures at 673K~1473K after using optimal parameters.Temperature measurement inaccuracy at 673K~1473K exceeds 0.3%.
  • 加载中
  • [1]

    KHAN M A,ALLEMAND C,EAGAR T W.Noncontact temperature measurement Ⅰ:interpolation based techniques[J].Rev Sci Instrum,1991,62(2):392~402.
    [2]

    HAHN J W,RHEE C.Reference wavelength method for a two-color pyrometer[J].Appl Opt,1987,26(24):5276~5279.
    [3]

    HUNTER G B,ALLEHAND C D,EAGER T W.Multiwavelength pyrometer[J].Opt Engng,1985,24(6):1081~1085.
    [4]

    DEWITT D P.Inferring temperature from optical radiation measurements[J].Opt Engng,1986,25(4):596~601.
    [5]

    KRIVTZUM V,GRASS B,HERGENRODER R et al.Temperature measurement of liquids by differential absorption of two diode laser[J].Applied Spectroscopy,2001,55(9):1251~1258.
    [6]

    FAN C H,LONGTIN J P.Laser-based measurement of temperature or concentration change at liquid surface[J].Journal of Heat Transfer,Transactions ASME,2000,122(4):757~762.
    [7]

    KEE H H,LEES G P,NEWSON T P.Technique for measuring distributed temperature with 35cm spatial resolution utilizing the Landau-Placzek ratio[J].IEEE Photon Technol Lett,2000,12(7):873~875.
    [8] 胡瑞华,栾松,孙宁克.可同时测量温度及发射率的辐射温度计[J].红外技术,1998,20(2):21~22.

    [9] 施德恒,黄国庆,刘玉芳.一种实用化实时测温系统的工作波长的优化设计[J].中国激光,2003,30(6):545~550.

    [10] 施德恒,黄国庆,刘玉芳 et al.一种实用化实时测温系统的优化设计[J].光学学报,2003;23(11):1394~1401.

    [11] 施德恒,黄国庆,刘玉芳.光源对测温系统测温精度及测温范围的影响[J].光电子*激光,2003,14(5):529~533.

    [12]

    TAMURA Y,TATSUWAKI M,SUGIMURA T et al.Temperature measure of steel in the furnace[A].Temperature,its Measurement and Control in Science and Industry[C].New York:American Institute of Physics,1982.505~512.
    [13] 王勇,廖延标,辛军 et al.实用化双折射式光纤温度传感器系统最佳设计[J].光学学报,1999,19(1):87~92.

    [14]

    COATES P B.Analytic estimation of systematic errors in photoelectric pyrometry[J].High Temperatures-High Pressures,1985,17:507~518.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article views(2555) PDF downloads(315) Cited by()

Proportional views

Temperature measurement sensitivity of a practical real-time temperature measurement system

  • 1. Department of Physics, Jinggangshan Normal College, Ji'an 343009, China;
  • 2. Department of Physics, Xinyang Teachers' College, Xinyang 464000, China;
  • 3. Department of Foundation, The First Aeronautical College of Air Force, Xinyang 464000, China;
  • 4. Department of Physics, Henan Normal University, Xinxiang 453002, China;
  • 5. Institute of Atomic and Molecular Physics, Sichuan University, Chengdu 610065, China

Abstract: Based on the Kirchhoff law,a practical temperature real-time measurement system using laser diode InGaAs/I and LiTaO3 pyroelectric detector is presented.By means of analyzing the relationships among resolving power of A/D converter,temperature measurement sensitivity of V(T)-T curve and temperature measurement range,16bite A/D converter for this instrument is determined.According to the relationships among temperature measurement accuracy,relative temperature measurement sensitivity of V(T)-T curve and waveleng ths,further optimal wavelength is discussed.Based on the relationships among resolving power of the detector and A/D converter and temperature sensitivity of V(T)-T curve,further optimal wavelength bandwidth is analyzed.Temperature measurement sensitivity of the instrument at 673K~1473K is given.Temperature measurement sensitivity is above 0.5K at lower temperatures and above 0.1K at higher temperatures at 673K~1473K after using optimal parameters.Temperature measurement inaccuracy at 673K~1473K exceeds 0.3%.

Reference (14)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return