Temperature measurement sensitivity of a practical real-time temperature measurement system
-
1.
Department of Physics, Jinggangshan Normal College, Ji'an 343009, China;
-
2.
Department of Physics, Xinyang Teachers' College, Xinyang 464000, China;
-
3.
Department of Foundation, The First Aeronautical College of Air Force, Xinyang 464000, China;
-
4.
Department of Physics, Henan Normal University, Xinxiang 453002, China;
-
5.
Institute of Atomic and Molecular Physics, Sichuan University, Chengdu 610065, China
-
Received Date:
2003-11-20
Accepted Date:
2004-04-17
-
Abstract
Based on the Kirchhoff law,a practical temperature real-time measurement system using laser diode InGaAs/I and LiTaO3 pyroelectric detector is presented.By means of analyzing the relationships among resolving power of A/D converter,temperature measurement sensitivity of V(T)-T curve and temperature measurement range,16bite A/D converter for this instrument is determined.According to the relationships among temperature measurement accuracy,relative temperature measurement sensitivity of V(T)-T curve and waveleng ths,further optimal wavelength is discussed.Based on the relationships among resolving power of the detector and A/D converter and temperature sensitivity of V(T)-T curve,further optimal wavelength bandwidth is analyzed.Temperature measurement sensitivity of the instrument at 673K~1473K is given.Temperature measurement sensitivity is above 0.5K at lower temperatures and above 0.1K at higher temperatures at 673K~1473K after using optimal parameters.Temperature measurement inaccuracy at 673K~1473K exceeds 0.3%.
-
-
References
[1]
|
KHAN M A,ALLEMAND C,EAGAR T W.Noncontact temperature measurement Ⅰ:interpolation based techniques[J].Rev Sci Instrum,1991,62(2):392~402. |
[2]
|
HAHN J W,RHEE C.Reference wavelength method for a two-color pyrometer[J].Appl Opt,1987,26(24):5276~5279. |
[3]
|
HUNTER G B,ALLEHAND C D,EAGER T W.Multiwavelength pyrometer[J].Opt Engng,1985,24(6):1081~1085. |
[4]
|
DEWITT D P.Inferring temperature from optical radiation measurements[J].Opt Engng,1986,25(4):596~601. |
[5]
|
KRIVTZUM V,GRASS B,HERGENRODER R et al.Temperature measurement of liquids by differential absorption of two diode laser[J].Applied Spectroscopy,2001,55(9):1251~1258. |
[6]
|
FAN C H,LONGTIN J P.Laser-based measurement of temperature or concentration change at liquid surface[J].Journal of Heat Transfer,Transactions ASME,2000,122(4):757~762. |
[7]
|
KEE H H,LEES G P,NEWSON T P.Technique for measuring distributed temperature with 35cm spatial resolution utilizing the Landau-Placzek ratio[J].IEEE Photon Technol Lett,2000,12(7):873~875. |
[8]
|
胡瑞华,栾松,孙宁克.可同时测量温度及发射率的辐射温度计[J].红外技术,1998,20(2):21~22.
|
[9]
|
施德恒,黄国庆,刘玉芳.一种实用化实时测温系统的工作波长的优化设计[J].中国激光,2003,30(6):545~550.
|
[10]
|
施德恒,黄国庆,刘玉芳 et al.一种实用化实时测温系统的优化设计[J].光学学报,2003;23(11):1394~1401.
|
[11]
|
施德恒,黄国庆,刘玉芳.光源对测温系统测温精度及测温范围的影响[J].光电子*激光,2003,14(5):529~533.
|
[12]
|
TAMURA Y,TATSUWAKI M,SUGIMURA T et al.Temperature measure of steel in the furnace[A].Temperature,its Measurement and Control in Science and Industry[C].New York:American Institute of Physics,1982.505~512. |
[13]
|
王勇,廖延标,辛军 et al.实用化双折射式光纤温度传感器系统最佳设计[J].光学学报,1999,19(1):87~92.
|
[14]
|
COATES P B.Analytic estimation of systematic errors in photoelectric pyrometry[J].High Temperatures-High Pressures,1985,17:507~518. |
-
-
Proportional views
-