Nonlinear fitting method based on dynamic calibration of thermocouples

  • Corresponding author: HAO Xiaojian, haoxiaojian@nuc.edu.cn
  • Received Date: 2013-07-01
    Accepted Date: 2013-08-22
  • In order to improve the accuracy of thermocouple dynamic calibration, based on a new thermocouple dynamic calibration system composed of a semiconductor laser, infrared detector and calibration thermocouples, the static calibration of the infrared detector was analyzed in the dynamic calibration system. According to the principle of back propagation(BP) neural network, the structure and the parameters of neural network were determined. The dynamic calibration experiment with ordinary K type armor-loaded thermocouple was performed and the static calibration data of infrared detector was acquired. Nonlinear curve fitting was performed using the least square method and BP neural network. The fitting results of the both the methods were analyzed and the fitting curves were obtained. The results show that, the fitting effect of BP neural network is better than the traditional least square method when there is less and uneven distributed data. The error caused by data fitting is reduced, dynamic characteristics of the thermocouple are acquired more accurately and thermocouple dynamic compensation is realized. The study has an important reference to the research of thermocouple dynamic characteristics.
  • 加载中
  • [1]

    LIANG L G,ZHOU H Ch,PAN D H. A study on the method of calibration for transient surface temperature detectors[J].Acta Armamentaria,2001,22(2):263-265(in Chinese).
    [2]

    QIAO L Sh,WANG Y L,ZENG J G. Discussion on methods of curve fitting in experimental data processing[J]. Journal of Chengdu University of Technology(Science & Technology Edition),2004,31(1):91-95(in Chinese).
    [3]

    HAO X J,LI K J,LIU J,et al.Traceability dynamic calibration of temperature sensor based on CO2 laser[J].Acta Armamentaria,2009,30(2):156-159(in Chinese).
    [4]

    HUANG L. Research on dynamic calibration technology of transient surface temperature sensor[D].Taiyuan: North University of China,2007:18-20(in Chinese).
    [5]

    HAO X J. Transient high surface temperature measurement and dynamic calibration[D]. Taiyuan:North University of China,2005: 18-20(in Chinese).
    [6]

    HE Q B.Back propagation neural network and applications[D].Chongqing: Chongqing Jiaotong University,2004:13-16(in Chinese).
    [7]

    TANG W M.The study of the optimal structure of BP nueral network[J]. Systems Engineering-theory & Practice,2005,25(10):96-100(in Chinese).
    [8]

    WU K, HARMS J. Performance study of a multipath routing method for wireless mobile ad hoc networks[C]// 9th International Symposium on Modeling, Analysis and Simulation of Computer and Telecommunication Systems.Cincinnati, Ohio, United States:IEEE, 2001: 99-107.
    [9]

    WU M X,ZHANG X L,WEN Sh H, et al.Summarization of BP neural networks improvement[J].Journal of Taiyuan University of Science and Technology,2005, 26(2):120-125(in Chinese).
    [10]

    HAO X J, LI K J,ZHOU H Ch, et al. Research on dynamic measurement errors compensation technology of thermocouple[J]. Journal of North University of China (Natural Science Edition), 2008,29(4):371-375(in Chinese).
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article views(3405) PDF downloads(615) Cited by()

Proportional views

Nonlinear fitting method based on dynamic calibration of thermocouples

    Corresponding author: HAO Xiaojian, haoxiaojian@nuc.edu.cn
  • 1. National Key Laboratory for Electronic Measurement Technology, North University of China, Taiyuan 030051, China;
  • 2. Key Laboratory of Instrumentation Science and Dynamic, North University of China, Taiyuan 030051, China

Abstract: In order to improve the accuracy of thermocouple dynamic calibration, based on a new thermocouple dynamic calibration system composed of a semiconductor laser, infrared detector and calibration thermocouples, the static calibration of the infrared detector was analyzed in the dynamic calibration system. According to the principle of back propagation(BP) neural network, the structure and the parameters of neural network were determined. The dynamic calibration experiment with ordinary K type armor-loaded thermocouple was performed and the static calibration data of infrared detector was acquired. Nonlinear curve fitting was performed using the least square method and BP neural network. The fitting results of the both the methods were analyzed and the fitting curves were obtained. The results show that, the fitting effect of BP neural network is better than the traditional least square method when there is less and uneven distributed data. The error caused by data fitting is reduced, dynamic characteristics of the thermocouple are acquired more accurately and thermocouple dynamic compensation is realized. The study has an important reference to the research of thermocouple dynamic characteristics.

Reference (10)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return